e-ISSN: 2394-2967
logo

British Journal of Medical and Health Research

Published

Procedure for calibration curve determination of radiochromic films for routine QA in Superficial Therapy X-Ray equipment

Published in June 2024 Issue 6 (Vol. 11, Issue 6, 2024)

Procedure for calibration curve determination of radiochromic films for routine QA in Superficial Therapy X-Ray equipment - Issue cover

Abstract

Purpose: This paper approaches the calibration procedure for EBT3 radiochromic films, and its precision, using a flat-bed scanner as a digitization system, for routine QA in a Superficial Therapy X-Ray machine. Methods: This paper analyzes precision in EBT3 radiochromic film calibration, using a flat-bed scanner as a digitization system, and describing routine control procedure within the framework of a quality assurance program for X-Ray equipment. Radiochromic films were cut into 5cm x 5cm squares, placed on 5cm of PMMA at 15cm source-film distance, and irradiated with several doses to obtain the calibration curve of net optical density (ODn) as a function of dose, and dose uncertainties were computed by residual analysis. Results: The calibration curve fits a fourth-degree polynomial function, with uncertainties of 2.10%, and 2.48% for the dispersion in the dose measurements in the films; we also considered a 3.5% uncertainty in dose measurements due to the use of an ionization chamber, for a net uncertainty (one sigma) of 4.77 % in the dose value. Conclusions: The calibration curve of the EBT3 films was obtained by a procedure that establishes the ODn as a function of the dose delivered by the X-Ray machine, based on the traceability of the dose.

Authors (3)

Anthony Cordero-Ramirez

Research Center in Materials S...

View all publications →

Walter Vargas-Segura

Physics Department. Costa Rica...

View all publications →

Esteban Avendaño-Soto

Research Center in Materials S...

View all publications →

Download Article

PDF

Best for printing and citation

File size: 0.0 MB
Format: PDF

Download Article

PDF

Best for printing and citation

File size: 0.0 MB
Format: PDF

Article Information

BJMHR1106002

BJMHR-11-000002

2024-06-01

Article Impact

Views:3,212
Downloads:1,171
scite_
PlumX Metrics Badge

How to Cite

Cordero-Ramirez & Vargas-Segura & Avendaño-Soto (2024). Procedure for calibration curve determination of radiochromic films for routine QA in Superficial Therapy X-Ray equipment. British Journal of Medical and Health Research, 11(6), xx-xx. DOI:https://doi.org/https://zenodo.org/records/13859865

Article Actions